YY311 water vapor transmission rate tester (infrared method), the instrument is suitable for the determination of water vapor transmission rate of plastic films, composite films and other films and sheet materials. Through the measurement of water vapor transmission rate, the technical indicators of control and adjustment of packaging materials and other products can be achieved to meet the different needs of product applications.
ASTM F1249、ISO 15106-2、TAPPI T557
1. Three chambers can simultaneously measure the water vapor transmission rate of the sample;
2. The three test chambers are completely independent and can test three identical or different samples at the same time;
3. Wide-range, high-precision temperature and humidity control, to meet the test under various test conditions;
4. The system adopts computer control, and the whole test process is completed automatically
5. Equipped with USB universal data interface to facilitate data transfer;
6. The software follows the principle of GMP authority management, and has functions such as user management, authority management, and data audit tracking.
The pretreated sample is clamped between the test chambers, nitrogen with a certain relative humidity flows on one side of the film, and dry nitrogen flows on the other side of the film. Due to the existence of the humidity gradient, water vapor will pass through the high humidity side. Diffusion through the film to the low-humidity side. On the low-humidity side, the permeated water vapor is carried to the sensor by the flowing dry nitrogen. When entering the infrared sensor, different spectral signals will be generated. Through the analysis and calculation of the different spectral signals, the sample is obtained. parameters such as water vapor transmission rate.
Test range: 0.01~40 g/(m2·24h)
Resolution: 0.01 g/m2 24h
Number of samples: 3 pieces (independently)
Sample size: 100mm×110mm
Test area: 50cm2
Sample thickness: ≤3mm
Temperature control range: 15℃~55℃
Temperature control accuracy: ±0.1℃
Humidity control range: 50%RH~90%RH;
Humidity control accuracy: ±2%RH
Carrier gas flow: 100 ml/min
Carrier gas type: 99.999% high purity nitrogen
Dimensions: 680×380×300 mm
Power supply: AC 220V 50Hz
Net weight: 72kg