IV.Test the principle
The principle of the moisture permeable cup weighing test is adopted. At a certain temperature, a specific humidity difference is formed on both sides of the sample. Water vapor passes through the sample in the moisture permeable cup and enters the dry side, and then is measured
The change in the weight of the moisture permeation cup over time can be used to calculate parameters such as the water vapor transmission rate of the sample.
V. Meeting the standard:
GB 1037、GB/T16928、ASTM E96、ASTM D1653、TAPPI T464、ISO 2528、YY/T0148-2017、DIN 53122-1、JIS Z0208、YBB 00092003、YY 0852-2011
VI.Product Parameters:标
Indicator |
Parameters |
Measure range |
Weight increase method:0.1 ~10 ,000g/㎡·24hWeight decrease method:0.1~2,500 g/m2·24h |
Sample qty |
3 The data are independent of each other.) |
Test accuracy |
0.01 g/m2·24h |
System resolution |
0.0001 g |
Temperature control range |
15℃ ~ 55℃(Standard)5℃-95℃(Can be custom-made) |
Temperature control accuracy |
±0.1℃(Standard) |
Humidity control range |
Weight loss method: 90%RH to 70%RHWeight gain method: 10%RH to 98%RH (The national standard requires 38℃ to 90%RH)
The definition of humidity refers to the relative humidity on both sides of the membrane. That is, for the weight loss method, it is the humidity of the test cup at 100%RH- the humidity of the test chamber at 10%RH-30%RH. The weight gain method involves the humidity of the test chamber (10%RH to 98%RH) minus the humidity of the test cup (0%RH) When the temperature varies, the humidity range changes as follows: (For the following humidity levels, the customer must provide a dry air source; otherwise, it will affect the generation of humidity.) Temperature: 15℃-40℃; Humidity: 10%RH-98%RH Temperature: 45℃, Humidity: 10%RH-90%RH Temperature: 50℃, Humidity: 10%RH-80%RH Temperature: 55℃, Humidity: 10%RH-70%RH |
Humidity control accuracy |
±1%RH |
Blowing wind speed |
0.5~2.5 m/s (Non-standard is optional) |
Sample thickness |
≤3 mm(Other thickness requirements can be customized 25.4mm ) |
Test area |
33 cm2(Options) |
Sample size |
Φ74 mm (Options) |
Volume of the test chamber |
45L |
Test mode |
The method of increasing or decreasing weight |
Gas source pressure |
0.6 MPa |
Interface size |
Φ6 mm (Polyurethane pipe) |
Power supply |
220VAC 50Hz |
External dimensions |
60 mm (L) × 480 mm (W) × 525 mm (H) |
Net weight |
70Kg |