| Model Name | YY2308B | |
| Standard | ISO13320-1:2009,GB/T19007-2016,Q/0100JWN001-2013Compliance with 21 CFR Part 11 | |
| Principle | Laser diffraction principle | |
| Analysis | Mie and Fraunhofer scattering | |
| Detector Arrangement | Log-spaced array, test angle from 0.015 degree to 145 degree | |
| Measuring Range | Wet:0.01μm-1200 μm Dry: 0.1μm-1200μm | |
| Silicon Photodetectors | Wet:127 pcs Dry:100 pcs | |
| Accuracy error | Wet<1% Dry<1% (CRM D50) | |
| Repeatability error | Wet<1% Dry<1% (CRM D50) | |
|
Light source |
High performance semiconductor red laser (λ=639nm )P>3.0MWAuxiliary green solid semiconductor laser (λ= 405 nm)P>2.0MW (available) | |
|
Optical path |
Converging light Fourier transform optical path | |
|
Effective focal length |
500mm | |
|
Laser Safety |
Class 1 | |
|
Wet dispersion |
Ultrasonic | Frequency:40KHz Power:60W, Time: ≥1S |
| Stir | Revolutions Speed: 0-3000RPM (Adjustable) | |
| Circulate | Rated Flow:30L/min Rated Power:70W | |
| Water level sensor (UK | Prevent water overflow and effectively protect the instrument | |
| Sample tank | Volume:1000mL | |
| Micro-Sample cuvette | Volume: 10mL ( Available) | |
| Dry dispersion | Dry-turbulence dispersion patent technology, normal shock wave shear technique | |
| Feeding Speed | Adjustable (Variable speed knob) | |
| Operation Mode | Full automatic / manual control, freely choose | |
| Dispersion medium | Compressed Air, pressure: 0 to 6 bar | |
| Optical bench alignment system | Full automatic, precision is up to 0.2um | |
| Full Test Speed per time | Wet: <2 Min Dry : <1minInverval time per test result :500ms | |
| Outer dimension | L104cm×W44cm×H54cm | |
| Net Weight | 70Kg | |