Technical Data
| Model | Basic Edition Haze Meter |
| Character | ASTM D1003/D1044 standard for haze and light transmittance measurement. Open measurement area and samples can be tested vertically and horizontally. Application: glass, plastic , film, display screen, packaging and other industries. |
| Illuminants | A,C |
| Standards | ASTM D1003/D1044, ISO13468/ISO14782, GB/T 2410,JJF 1303-2011, CIE 15.2, GB/T 3978, ASTM E308, JIS K7105, JIS K7361, JIS K 7136 |
| Test Parameter | ASTM (HAZE),Transmittance(T) |
| Test Aperture | 21mm |
| Instrument Screen | 5 inches color LCD screen |
| Haze Repeatability | Φ21mm aperture, Standard Deviation: within 0.1(when a haze standard with value 40 is measured 30 times at 5-second interval after calibration) |
| Transmittance Repeatability | ≤0.1 unit |
| Geometry | Transmittance 0/D (0 degree illumination, diffused receiving) |
| Integrating Sphere Size | Φ154mm |
| Light Source | 400~700nm full spectrum LED light source |
| Test Range | 0-100% |
| Haze Resolution | 0.01 unit |
| Transmittance Resolution | 0.01 unit |
| Sample Size | Open space, no size limit |
| Data Storage | 10,000 pcs of samples |
| Interface | USB |
| Power Supply | DC12V (110-240V) |
| Working Temperature | +10 – 40 °C (+50 – 104 °F) |
| Storage Temperature | 0 – 50 °C (+32 – 122 °F) |
| Instrument Size | L x W x H: 310mmX215mmX540mm |